16-2 :CONFigure Subsystem NBFM Commands
16-2 PN: 10580-00319 Rev. N S412E PM
16-2 :CONFigure Subsystem
This set of commands prepares the instrument for the selected measurement. It disables any
currently-enabled measurements and activates the specified measurement.
Current instrument settings may be changed to default values. These changes are identified
with their respective measurement commands.
:CONFigure?
Description: :CONFigure? query returns the name of the measurement previously
set up using a CONFigure command or a MEASure? query. The list
below shows the possible return values and the actual names of each
configuration.
Cmd Parameters: NA
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To query the current measurement type:
:CONFigure?
Front Panel Access: Measurement
Note
Sending a non-query :CONFigure command will change the Sweep setting from
Run to Hold.
Returns Value Actual Name
SIGA NBFM Analyzer
COV NBFM Coverage
QUI NBFM Quieting
SNDR NBFM SINAD