5-11 [:SENSe] Subsystem VNA Commands
5-80 PN: 10580-00319 Rev. N S412E PM
[:SENSe]:SWEep:TYPE SINGle|CONTinuous
[:SENSe]:SWEep:TYPE?
Description: Sets the sweep type. The query version of this command returns
“SING” if current sweep is set to single sweep, and returns “CONT” if
set to continuous sweep. Note that setting the sweep type to SINGle
sets the sweep to hold.
Cmd Parameter: <char> SINGle|CONTinuous
Query Response: <char> SING|CONT
Default Value: CONT
Front Panel Access: Shift-3 (Sweep), Sweep Type
[:SENSe]:TRACe Subsystem
This subsystem includes commands that provide general settings for each trace.
[:SENSe]:TRACe<Tr>:DOMain FREQuency|DISTance
[:SENSe]:TRACe<Tr>:DOMain?
Description: Defines the domain for the given trace <Tr>. <Tr> is the trace
number in the range 1 to 4. If no trace number is specified, then the
<Tr> parameter defaults to trace number 1. The query version of this
command returns “FREQ” if domain is Frequency, and “DIST” if
domain is distance.
Syntax: :TRACe<Tr>:DOMain FREQuency|DISTance
:TRACe<Tr>:DOMain?
Cmd Parameter: <char> FREQuency|DISTance
Query Response: <char> FREQ|DIST
Default Value: Trace 1: FREQ
Trace 2: FREQ
Trace 3: FREQ
Trace 4: FREQ
Front Panel Access: Measure, Domain Selection