VNA Commands 5-11 [:SENSe] Subsystem
S412E PM PN: 10580-00319 Rev. N 5-81
[:SENSe]:TRACe<Tr>:SELect
Description: Selects the given trace, <Tr>, as the active trace. <Tr> is the trace
number in the range 1 to 4. If no trace number is specified, then the
<Tr> parameter defaults to trace number 1. Note that this may also
change the total number of traces.
Cmd Parameter: NA
Query Response: NA (no query)
Default Value: TR1
Example: To set trace 2 as the active trace:
:SENSe:TRACe2:SELect
:SENS:TRAC2:SEL
To set trace 1 as the active trace:
:SENSe:TRACe1:SELect
:SENSe:TRACe:SELect
Front Panel Access: Measure, Active Trace
[:SENSe]:TRACe<Tr>:SPARams S11|S21
[:SENSe]:TRACe<Tr>:SPARams?
Description: Defines the S-parameter for the given trace, <Tr>.
<Tr> is the trace number in the range 1 to 4. If no trace number is
specified, then the <Tr> parameter defaults to trace number 1. The
query version of this command returns “S11” if the S-parameter is set
to S11, “S21” if set to S21.
Syntax: :TRACe<Tr>:SPARams S11|S21
:TRACe<Tr>:SPARams?
Cmd Parameter: <char> [S11|S21
Query Response: <char> [S11|S21
Default Value: Trace 1: S11
Trace 2: S21
Trace 3: S11
Trace 4: S21
Example: To assign S11 to trace 2:
:SENSe:TRACe2:SPARams S11
Front Panel Access: Measure, S Parameter