EasyManua.ls Logo

Anritsu LMR Master S412E - 15-3:DISPlay Subsystem

Anritsu LMR Master S412E
704 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15-3 :DISPlay Subsystem PTC Commands
15-4 PN: 10580-00319 Rev. N S412E PM
15-3 :DISPlay Subsystem
This subsystem provides commands that modify the display of data for the user. They do not
modify the way in which data are returned to the controller.
:DISPlay[:WINDow]:TRACe:SELect?
Description: This command returns the current active trace number in the format
TRAC#.
Cmd Parameters: NA
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To query for the active trace number:
:DISPlay:TRACe:SELect?
Front Panel Access: Measurement, PTC Analyzer, Active Graph
:DISPlay[:WINDow]:TRACe:Y[:SCALe]:PDIVision <value>
:DISPlay[:WINDow]:TRACe:Y[:SCALe]:PDIVision?
Description: Sets the scale per division for the y-axis. In the PTC Analyzer
measurement, this value corresponds to the scale on the spectrum
graph type.
Cmd Parameters: <value>
Query Parameters: NA
Range: 1 to 15
Default Value: 10
Default Unit: NA
Example: To set the scale to 8:
:DISPlay:TRACe:Y:PDIVision 8
Front Panel Access: Amplitude, Scale

Table of Contents