PTC Commands 15-4 :FETCh Subsystem
S412E PM PN: 10580-00319 Rev. N 15-7
15-4 :FETCh Subsystem
This set of commands returns the most recent measurement data of the active measurement.
They will not switch to another measurement.
To prepare for a new measurement, use the CONFigure command. To make a new
measurement, use the INITiate command. To get new measurement data, use the READ or
MEASure query commands.
:FETCh:COVerage?
Description: Returns the most recent PTC Coverage numerical measurement
results. Data is returned as 7 comma-separated values in the following
order and format:
RSSI (dBm as float)
BER (% as float)
Mod Fid (% as float)
Latitude (radian as float)
Longitude (radian as float)
UTC Date (month/day/year as char)
UTC Time (hour:minute:second as char)
If there is no valid measurement data, the instrument will return
“--,--,--,--,--,--,--”.
Cmd Parameters: NA
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To fetch PTC Coverage numerical data:
:FETCh:COVerage?
Related Command: :CONFigure:COVerage
Front Panel Access: NA