VNA Commands 5-11 [:SENSe] Subsystem
S412E PM PN: 10580-00319 Rev. N 5-79
[:SENSe]:SWEep Subsystem
This subsystem includes commands that affect the sweep parameters of the instrument.
[:SENSe]:SWEep:IFBW <freq value>
[:SENSe]:SWEep:IFBW?
Description: Sets the IF Bandwidth. The <freq value> in Hz must be one of the
following 13 values:
100000|50000|20000|10000|5000|2000|1000|500|200|
100|50|20|10
The query form of this command returns the frequency in Hz.
Cmd Parameter: <char> <freq value>
Query Response: <char> <freq value>
Default Value: 10000
Default Units: Hz
Range: 10 to 100000 Hz
Example: Sets the IF Bandwidth frequency to 200 Hz:
:SENS:SWE:IFBW 200
Sets the IF Bandwidth frequency to 100 kHz:
:SENS:SWE:IFBW 100000
Front Panel Access: Shift-3 (Sweep), IFBW
[:SENSe]:SWEep:POINts <integer>
[:SENSe]:SWEep:POINts?
Description: Sets the total number of measurement points per sweep. Note that a
sweep with a lower number of data points will complete in less time
than a sweep with a higher number of data points.
Syntax: :SWEep:POINts <integer>
:SWEep:POINts?
Cmd Parameter: <NR1> <integer>
Query Response: <NR1> <integer>
Default Value: 201
Range: 2to4001
Front Panel Access: Shift-3 (Sweep), Data Points