5-11 [:SENSe] Subsystem VNA Commands
5-78 PN: 10580-00319 Rev. N S412E PM
[:SENSe]:FREQuency:STOP
Description: Sets the stop frequency. Note that changing the value of the stop
frequency will change the value of coupled
parameters: Center Frequency and Span.
Syntax: :FREQuency:STOP <freq>
:FREQuency:STOP?
Cmd Parameter: <NRf> <freq> (hertz)
Query Response: <NR3> <freq> (hertz)
Default Value: For 1.6 GHz Model: 1600000000 Hz
For 6 GHz Model: 6000000000 Hz
Default Units: Hz
Range: For 1.6 GHz Model: 500 kHz to 1.6 GHz
For 6 GHz Model: 500 kHz to 6 GHz
Example: Sets the stop frequency to 10000 Hz:
:SENSe:FREQuency:STOP 10000
Sets the stop frequency to 5 MHz:
:SENSe:FREQuency:STOP 5MHZ
Sets the stop frequency to 1 GHz:
:SENS:FREQ:STOP 1GHZ
Front Panel Access: Freq/Dist (or Freq), Stop Freq
:SENSe<Tr>:FREQuency:DATA?
Description: Produces the frequency list in Hz for the given trace. <Tr> is the trace
number in the range 1 to 8 (1 to 4 for Traces TR1 to TR4 and 5 to 8 for
Memory M1 to M4). If no trace number is specified, then the <Tr>
parameter defaults to trace number 1. The response begins with an
ASCII header that specifies the number of data bytes. It appears in the
format #AX, where A is the number of digits in X, and X is the
number of bytes that follow the header. Each frequency point is in
scientific notation and separated by a comma delimiter.
Cmd Parameter: NA (query only)
Query Response: <char> <freq> (returns block data in hertz)
Default Units: Hz
Related Command: :CALCulate<Tr>:TRANsform:DISTance:DATA?
Front Panel Access: NA