VNA Commands 5-2 :CALCulate Subsystem
S412E PM PN: 10580-00319 Rev. N 5-23
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:DELTa[:STATe] OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:DELTa[:STATe]?
Description: Sets the specified marker as the active marker and turn it on or off. If
the value is set to ON or 1, then the specified marker is turn on and set
as a delta marker. If the value is set to OFF or 0, then the specified
marker is turn off. The query version of the command returns a 1 if the
specified marker is a delta marker, and returns a 0 if it is not a delta
marker.
Cmd Parameter: <boolean> OFF|ON|0|1
Query Response: <bNR1> 0|1
Default Value: OFF
Example: To turn on marker #3 and set it as a delta marker:
:CALCulate:MARKer3:DELTa ON
:CALCulate:MARKer3:DELTa 1
:CALCulate:MARKer3:DELTa:STATe ON
:CALCulate:MARKer3:DELTa:STATe 1
To turn off delta marker #6:
:CALCulate:MARKer6:DELTa OFF
:CALCulate:MARKer6:DELTa:STATe OFF
:CALCulate:MARKer6:DELTa:STATe 0
Related Command: :CALCulate:MARKer:DELTa:REFerence
Front Panel Access: Marker, Maker Type
:CALCulate:MARKer:DISPlay:FORMat NONE|SCREen|TABLe|TRACe
:CALCulate:MARKer:DISPlay:FORMat?
Description: Sets the display readout format for markers. The query version of the
command returns “NONE” if the display readout format is set to None,
“SCRE” if Screen, “TABL” if Table, and “TRAC” if Trace.
Cmd Parameter: <char> NONE|SCREen|TABLe|TRACe
Query Response: <char> NONE|SCRE|TABL|TRAC
Default Value: NONE
Example: To set marker readout format to Table:
:CALCulate:MARKer:DISPlay:FORMat TABLe
:CALCulate:MARKer:DISPlay:FORMat TABL
:CALC:MARK:DISP:FORM TABL
Front Panel Access: Marker, Readout Format