7-11 [:SENSe] Subsystem Fixed WiMAX Commands
7-28 PN: 10580-00319 Rev. N S412E PM
[:SENSe]:DEMod:SFLatness:STOP:X?
Title: Spectral Flatness Stop Value
Description: Queries the Spectral Flatness stop value.
Related Command: :DEMod:SFLatness:STARt:X?
[:SENSe]:DLFLength 2.5|5|10
[:SENSe]:DLFLength?
Title: Down Link Frame Length
Description: Sets the Down Link Frame Length, in milliseconds.
The query form of this command will return the Down Link Frame
Length in seconds (not milliseconds).
Parameter: 2.5|5|10
Default Value: 2500 us
Default Unit: seconds
Front Panel Access: Setup, Frame Length
[:SENSe]:FREQuency:CENTer <freq>
[:SENSe]:FREQuency:CENTer?
Title: Center Frequency
Description: Sets the center frequency. Note that changing the value of the center
frequency will change the value of the coupled parameters Start
Frequency and Stop Frequency.
Parameter: <freq>
Default Value: 2.5 GHz
Default Unit: Hz
Range: 0 Hz to 7.1 GHz
Front Panel Access: Freq, Center Freq
[:SENSe]:FREQuency:SIGStandard:CHANnel <number>
[:SENSe]:FREQuency:SIGStandard:CHANnel?
Title: Channel Selection
Description: Sets the channel number for the selected signal standard.
Parameter: <number>
Front Panel Access: Freq, Channel