Mobile WiMAX Commands 8-6 :INITiate Subsystem
S412E PM PN: 10580-00319 Rev. N 8-11
8-6 :INITiate Subsystem
This subsystem controls the triggering of measurements.
:INITiate:CONTinuous OFF|ON|0|1
:INITiate:CONTinuous?
Title: Continuous/Single Sweep
Description: Specifies whether the sweep/measurement is triggered continuously. If
the value is set to ON or 1, another sweep/measurement is triggered as
soon as the current one completes. If continuous is set to OFF or 0, the
instrument enters the “idle” state and waits for the
:INITiate[:IMMediate] command or for :INITiate:CONTinuous ON. The
default value is ON. That is, sending :INIT:CONT is equivalent to
sending :INIT:CONT ON. The query version of the command returns a
1 if the instrument is continuously sweeping/measuring and returns a 0
if the instrument is in single sweep/measurement mode. Note that rapid
toggling between ON and OFF is not allowed. The instrument must be
allowed to make a full sweep before toggling can be done.
Parameter: OFF|ON|0|1
Parameter Type: <boolean>
Default Value: ON
Related Command: :INITiate[:IMMediate]
:INITiate:HOLD
Front Panel Access: Shift-3 (Sweep), Sweep
:INITiate[:IMMediate]
Title: Trigger Sweep/Measurement
Description: Initiates a sweep/measurement. If :INITiate:CONTinuous is set to ON,
this command is ignored. Use this command in combination with
:STATus:OPERation? to synchronize the capture of one complete set of
data. When this command is sent, the “sweep complete” bit of
:STATus:OPERation? is set to 0, indicating that the measurement has
not completed. The data collection is then triggered. The controlling
program can poll :STATus:OPERation? to determine the status. When
the “sweep complete” bit is set to 1, data is ready to be retrieved.
Related Command: :INITiate:CONTinuous
:STATus:OPERation?
Front Panel Access: Shift-3 (Sweep), Trigger Sweep