9-8 :MEASure Subsystem LTE Commands
9-16 PN: 10580-00319 Rev. N S412E PM
:MEASure:RF:ACLR?
Title:
Measure Adjacent Channel Leakage Ratio
Description:
Sets the active measurement to adjacent channel power ratio, sets the
default measurement parameters, triggers a new measurement and
returns the main channel(s) power, and adjacent channel(s) power. It is
a combination of the commands :CONFigure:RF ACLR and
:READ:RF:ACLR? Refer to “:READ:RF:ACLR?” on page 9-19 for a
description of the returned data.
Default Unit:
dBm
Front Panel Access:
Measurements, RF, ACLR
:MEASure:RF:SEM?
Title:
Measure Spectral Emission Mask
Description:
Sets the active measurement to Spectral Emission Mask and returns a
PASS or FAIL result.
Front Panel Access:
Measurements, RF, Spectral Emission Mask
:MEASure:RF:SPECtrum?
Title:
Measure RF Spectrum
Description:
Sets the active measurement to Channel Spectrum, sets the default
measurement parameters, triggers a new measurement and returns the
Channel Power and Occupied Bandwidth measurement results. It is a
combination of the following command :CONFigure:RF SPECtrum and
:READ:RF:SPECtrum? Data is returned as 2 comma-delimited values:
Channel Power in dBm and Occupied bandwidth in MHz.
Front Panel Access:
Measurements, RF, Channel Spectrum