10-8 :MMEMory Subsystem P25 Phase 1 Commands
10-18 PN: 10580-00319 Rev. N S412E PM
:MMEMory:STORe:STATe <integer>,<filename>
Description: Stores the current setup into the file specified by <filename>.
<filename> should be enclosed in either single quotes (' ') or double
quotes (" ") and should not contain a file extension. Use the command
MMEMory:MSIS to set the current save location. The <integer>
parameter is not currently used, but it must be sent. Send a value of 0.
Cmd Parameters: <integer>,<filename>
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To save a setup file:
:MMEMory:STORe:STATe 0,"xxx"
Front Panel Access: Shift 7 (File), Save
:MMEMory:STORe:TRACe <integer>,<filename>
Description: Stores the trace into the file specified by <filename>. <filename>
should be enclosed in either single quotes (' ') or double quotes (" ") and
should not contain a file extension. Use the command MMEMory:MSIS
to set the current save location. The <integer> parameter is not
currently used, but it must be sent. Send a 0.
Cmd Parameters: <integer>,<filename>
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To save a measurement file:
:MMEMory:STORe:TRACe 0,"xxx"
Front Panel Access: Shift 7 (File), Save Measurement
Note
Control Channel, Bit Capture and IQ Data measurements can not be recalled on
the instrument.
Note
Control Channel, Bit Capture and IQ Data measurements can not be saved on the
instrument.