3-4 :MMEMory Subsystem All Mode Commands
3-6 PN: 10580-00319 Rev. N S412E PM
“.lte” for LTE Analyzer measurements
“.p25” for P25 Analyzer measurements
“.p252” for P25p2 Analyzer measurements
“.nxdn” for NXDN Analyzer measurements
“.dpmr” for dPMR Analyzer measurements
“.dmr2” for DMR 2 Analyzer measurements
“.ptc” for PTC Analyzer measurements
“.nbfm” for NBFM Analyzer measurements
“.tetra” for TETRA Analyzer measurements
Use the command MMEMory:MSIS to set the save location.
Cmd Parameter: <file name>
Query Response: NA (no query)
Front Panel Access: Shift-7 (File), Delete, Select or De-Select, Delete
:MMEMory:MSIS INTernal|USB
:MMEMory:MSIS?
Title: Save Location
Description: Sets the instrument’s internal memory or the USB Flash drive as the
save location for all subsequently saved files. This command also
determines the destination location for copied files. For example,
selecting internal memory as the current save location will set the USB
Flash drive as the destination for copied files, and vice-versa.
Note that the save location specified here applies to remote operation. It
is independent of and can be different from the save location set via the
instrument front panel. The query form of this command returns the
save location setting for remote operation, not the front panel setting.
Commands to load, store (save), or copy data will fail if the save location
selected is the USB drive and no USB device is plugged into the
instrument.
Before setting the save location, send the :SYSTem:MSIS? USB
command to query the ready state of the USB Flash drive.
Cmd Parameter: INTernal|USB
Query Response: INT|USB
Parameter Type: <char>
Related Command: :MMEMory:MSIS:DESTination
:SYSTem:MSIS[:STATe]?
Front Panel Access: Shift-7 (File), Save, Change Save Location, (select drive or folder)