DMR 2 Commands 14-3 :DISPlay Subsystem
S412E PM PN: 10580-00319 Rev. N 14-7
:DISPlay[:WINDow]:TRACe<Tr>:FORMat:SIGAnalyzer <graph type>
:DISPlay[:WINDow]:TRACe<Tr>:FORMat:SIGAnalyzer?
Description: Defines the graph type for the given trace <Tr>. <Tr> is the trace
number in the range 1 to 4. If no trace number is specified, then the
<Tr> parameter defaults to trace number 1. <Graph Type> is the
graph type to which the specified trace is set, and it must be one of the
following values:
CONStellation|SPECtrum|HISTogram|SUMMary|DMR2summary|
EYEDiagram|LINConstellation|PROFile
The query version of this command returns CONS” if the specified trace
graph type is set to Constellation, “HIST” if set to Histogram, “SPEC” if
set to Spectrum, “SUMM” if set to Summary, “DMR2” if set to
DMR 2 Summary, “EYED” if set to Eye Diagram, “LINC” if set to linear
constellation, and “PROF” if set to power profile.
Please note that this command only works when the current
measurement is set to DMR 2 Analyzer.
Cmd Parameters: <graph type>
Query Parameters: NA
Range: CONStellation|SPECtrum|HISTogram|SUMMary|DMR2summary|
EYEDiagram| LINConstellation|PROFile
Default Value: Trace 1: Linear Constellation
Trace 2: Spectrum
Trace 3: Histogram
Trace 4: Summary
Default Unit: NA
Example: To set Trace 2 graph type to Eye Diagram:
:DISPlay:TRACe2:FORMat:SIGAnalyzer EYEDiagram
Related Command: :CONFigure:SIGAnalyzer
Front Panel Access: Measurement, DMR 2 Analyzer, Graph Type