EasyManua.ls Logo

Anritsu LMR Master S412E - Page 478

Anritsu LMR Master S412E
704 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14-8 :MMEMory Subsystem DMR 2 Commands
14-22 PN: 10580-00319 Rev. N S412E PM
:MMEMory:STORe:TRACe <integer>,<filename>
Description: Stores the trace into the file specified by <filename>. <filename>
should be enclosed in either single quotes (' ') or double quotes (" ") and
should not contain a file extension. Use the command MMEMory:MSIS
to set the current save location. The <integer> parameter is not
currently used, but it must be sent. Send a 0.
Cmd Parameters: <integer>,<filename>
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To save a measurement file:
:MMEMory:STORe:TRACe 0,"xxx"
Front Panel Access: Shift 7 (File), Save Measurement
Note
Control Channel, Bit Capture and IQ Data measurements can not be saved on the
instrument.

Table of Contents

Related product manuals