14-13 [:SENSe] Subsystem DMR 2 Commands
14-48 PN: 10580-00319 Rev. N S412E PM
[:SENSe]:SYMBolspan <value>
[:SENSe]:SYMBolspan?
Description: Sets the symbol span. Please note that this setting only affects the
Eye Diagram in the DMR 2 Analyzer measurement. Please note that
setting the symbol span will restart the sweep.
Cmd Parameters: <value>
Query Parameters: NA
Range: 2 to 5
Default Value: 2
Default Unit: NA
Example: To set the symbol span to 4:
:SYMBolspan 4
Front Panel Access: Measurement, DMR 2 Analyzer, Symbol Span