PTC Commands 15-7 :MEASure Subsystem
S412E PM PN: 10580-00319 Rev. N 15-15
Note: The squelch setting [:SENSe]:DM:SQUelch will blank out (--)
all summary measurements on the instrument display except for
Received Pwr when the received power level is lower than the squelch
power setting. The received power level is also affected by the Rx Power
Offset setting. The query command will still return values even if the
instrument display is blanked out.
Cmd Parameters: NA
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To measure PTC Analyzer numerical data:
:MEASure:SIGAnalyzer?
Front Panel Access: NA