NBFM Commands 16-9 :READ Subsystem
S412E PM PN: 10580-00319 Rev. N 16-27
:READ:SNDRatio?
Description: Triggers a new NBFM SINAD measurement and returns the numerical
results. It is a combination of the commands :ABORT, :INITiate, and
:FETCh:SNDRatio? NBFM SINAD must be the active measurement
(specified by :CONFigure:SNDRatio). The current measurement can
be queried using :CONFigure?
Data is returned as 1 value in the following order and format:
Measured SINAD (dB as float).
If there is no valid measurement data, the instrument will return “--”.
Cmd Parameters: NA
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To read NBFM SINAD numerical data:
:READ:SNDRatio?
Front Panel Access: NA