16-13 [:SENSe] Subsystem NBFM Commands
16-42 PN: 10580-00319 Rev. N S412E PM
[:SENSe]:AVERage:COUNt <integer>
[:SENSe]:AVERage:COUNt?
Description: Sets the number of times the deviation value in the NBFM Analyzer
Summary window is averaged.
Cmd Parameters: <integer>
Query Parameters: NA
Range: 1 to 25
Default Value: 1
Default Unit: NA
Example: To set averaging to 15:
:AVERage:COUNt 15
Front Panel Access: Setup, Averaging
[:SENSe]:BANDwidth[:RESolution] 5|6.25|10|12.5|30|50
[:SENSe]:BANDwidth[:RESolution]?
[:SENSe]:BWIDth[:RESolution] 5|6.25|10|12.5|30|50
[:SENSe]:BWIDth[:RESolution]?
Description: Sets the IF bandwidth.
Note: IF bandwidth value is set and returned in kHz.
Cmd Parameters: 5|6.25|10|12.5|30|50
Query Parameters: NA
Range: 5|6.25|10|12.5|30|50
Default Value: 10
Default Unit: kHz
Example: To change the IF Bandwidth to 50 kHz:
:BANDwidth 50
:BWIDth 50
Front Panel Access: Setup, Filters, IFBW