17-3 :DISPlay Subsystem TETRA Commands
17-6 PN: 10580-00319 Rev. N S412E PM
:DISPlay[:WINDow]:TRACe<Tr>:FORMat:SIGAnalyzer <graph type>
:DISPlay[:WINDow]:TRACe<Tr>:FORMat:SIGAnalyzer?
Description: Defines the graph type for the given trace <Tr>. <Tr> is the trace
number in the range 1 to 4. If no trace number is specified, then the
<Tr> parameter defaults to trace number 1. <Graph Type> is the
graph type to which the specified trace is set, and it must be one of the
following values:
CONStellation|SPECtrum|SUMMary|EYEDiagram|TETRasummary
The query version of this command returns “CONS” if the specified trace
graph type is set to Constellation, “SPEC” if set to Spectrum, “SUMM” if
set to Summary, “EYED” if set to Eye Diagram, and “TETR” if set to
TETRA Summary.
Please note that this command only works when the current
measurement is set to TETRA Analyzer.
Cmd Parameters: <graph type>
Query Parameters: NA
Range: CONStellation|SPECtrum|SUMMary|EYEDiagram|TETRasummary
Default Value: Trace 1: Constellation
Trace 2: Spectrum
Trace 3: Tetra Summary
Trace 4: Summary
Default Unit: NA
Example: To set Trace 2 graph type to Eye Diagram:
:DISPlay:TRACe2:FORMat:SIGAnalyzer EYEDiagram
Related Command: :CONFigure:SIGAnalyzer
Front Panel Access: Measurement, TETRA Analyzer, Graph Type
:DISPlay[:WINDow]:TRACe<Tr>:SELect
Description: Selects the given trace, <Tr>, as the active trace. <Tr> is the trace
number in the range 1 to 4 for TETRA Analyzer. If no trace number is
specified, then the <Tr> parameter defaults to trace number 1.
Cmd Parameters: NA
Query Parameters: NA
Range: NA
Default Value: TRAC1
Default Unit: NA
Example: To set trace 2 as the active trace:
:DISPlay:TRACe2:SELect
Front Panel Access: Measurement, TETRA Analyzer, Active Graph