What to do if ASP Laboratory Equipment says 'Cassette Expired' or 'Cassette Out of Date'?
J
Jared JacobsonSep 1, 2025
If the ASP Laboratory Equipment indicates that the cassette has expired, is out of date, or has already been used, load a new cassette. If the message persists, run diagnostics. If diagnostics fail, adjust or replace the delivery subsystem.
B
biancafordSep 12, 2025
What to do if ASP Laboratory Equipment displays 'Delivery System Not Ready'?
T
thowardSep 12, 2025
If the ASP Laboratory Equipment displays 'Delivery System Not Ready', run diagnostics. Replace the failed component.
R
Roberto SimmonsAug 30, 2025
What to do if ASP sterrad NX shows 'H2O2 Adjustment Failed' or related H2O2 errors?
C
Christopher BradyAug 31, 2025
If the ASP Laboratory Equipment displays an H2O2 adjustment failure, bulb/detector fault, monitoring failure, or sensor fault, verify the UV lamp is working and replace it if necessary. Clean the optical windows on the lamp and detector. Run diagnostics and replace any failed components.
B
Brandon BellSep 14, 2025
What does 'H2O2 Rate Constant Too High' mean on ASP sterrad NX Laboratory Equipment?
B
Brian RichardsonSep 14, 2025
If the ASP Laboratory Equipment displays 'H2O2 Rate Constant Too High', it means the calculated rate constant for the hydrogen peroxide is greater than the threshold value. Run diagnostics and replace any failed components.
S
Steven GillSep 13, 2025
What to do if ASP Laboratory Equipment displays 'High Plasma Power'?
M
millermichaelSep 14, 2025
If the ASP Laboratory Equipment displays 'High Plasma Power', run diagnostics. If the diagnostics fail, replace the LFPS II power supply.
G
Garrett WashingtonSep 18, 2025
What to do if ASP sterrad NX displays 'Unable to Open One Second Data File'?
S
Steven HughesSep 18, 2025
If the ASP Laboratory Equipment displays 'Unable to Open One Second Data File', install a new PCMCIA compact flash memory card.
M
Melody LopezSep 23, 2025
What to do if ASP sterrad NX Laboratory Equipment displays 'Unknown Reason'?
D
Dustin StephensSep 23, 2025
If the ASP Laboratory Equipment displays 'Unknown Reason', start cycle again. If problem persists, run diagnostics.
S
Stanley HoganAug 27, 2025
What to do if ASP sterrad NX Laboratory Equipment displays 'Condenser Temperature' error?
A
Alyssa HamiltonAug 27, 2025
If the ASP Laboratory Equipment shows an error related to the condenser temperature (has not risen, too high, or too low), run diagnostics to see if a condenser heater or sensor has failed. You may need to adjust or replace the vaporizer/condenser assembly.
Z
Zachary LiAug 26, 2025
What to do if ASP sterrad NX displays 'Door Opened' error?
L
Lori ReedAug 26, 2025
If the ASP Laboratory Equipment detects that the door was opened during a cycle, and the problem persists, run diagnostics. You may need to troubleshoot or replace the door sensor.
J
Joshua MckeeAug 25, 2025
What to do if ASP sterrad NX Laboratory Equipment cannot eject the cassette?
J
jknoxAug 26, 2025
If the ASP Laboratory Equipment cannot eject the cassette, run diagnostics. If diagnostics fail, adjust or replace the delivery subsystem.