Ch. 10. Environmental Scanner Overview Sec. 10.3. Cypher ES Quick Reminder List
4.
Warning: Be careful to avoid touching the fragile piezo wires.
Hold the cantilever holder only by its handles.
Hold the cantilever holder only by its two plastic handles (upper image).
In particular, avoid contacting the wires running to the tapping piezo (lower image).
Be careful not to scrape the piezo wires when tightening the right side cantilever
holder locking screw.
Warning: Pay attention! The piezo wiring is very fragile. Be careful not to touch it.
5.
Always power off the microscope before
exchanging scanners.
Before exchanging scanners, remember
to first motor the objective to its upper
limit of travel.
Then power off the controller before
disconnecting the scanner.
Exchange scanners and then power the
controller back on after the new scanner
has been reconnected.
Warning: Pay attention! On older Cyphers,
exchanging the scanner while the system is
powered up can damage the backpack
electronics.
DRAFT
Page 124