Ch. 10. Environmental Scanner Overview Sec. 10.3. Cypher ES Quick Reminder List
10.3.2. Reminders for cell exchange and handling
1.
The fitting wrench should never be used to
attach gas lines to the cell body.
The fitting wrench (114.800) supplied in
the accessory kit should NEVER be used
with the cell bodies (see top right
image).
The fitting wrench should only be used
with the fitting fixture (114.721), which
is also supplied in the accessory kit, or
the manifold on the front of the scanner.
See the bottom right image.
Warning: Pay attention! Using the fitting
wrench on the cell body will crack the cell
body glass.
2.
Be careful not to poke the sample stage
diaphragms.
Be careful never to poke the sample
stage diaphragms with tweezers or
screwdrivers as this may cause a hole in
the diaphragm.
Some versions of the sample stages
have permanently attached diaphragms.
These stages cannot be repaired if their
diaphragms are damaged.
Warning: Pay attention! Don’t poke a hole
in your sample diaphragm by being careless.
DRAFT
Page 126