Ch. 4. Tutorial: AC Mode in Air, Std. Scanner Sec. 4.2. Loading Lever and Sample
11.
Tighten the clamp:
Once the cantilever holder is fully
inserted, use the ball head wrench to
gently tighten the clamp.
12.
Remove the old cantilever:
Position the changing station as shown,
on a flat hard surface.
Take the tweezers in your dominant
hand.
Press down on the station, as shown,
with your other hand. This depresses a
button on the bottom of the station
which drives a pin up under the
cantilever retaining clip.
Remove the cantilever and release
pressure on the station.
Inspect the cantilever area for tiny
silicon grit and blow clean with
compressed air if necessary.
13.
Select new cantilever:
Select a new cantilever and pick it up
with tweezers.
Close the box! Ruining $1k of levers by
putting your hand on an open box is not
unheard of.
Note If your lab saves some old cantilevers,
consider practicing with a “dummy”
cantilever.
Tip Some find it useful to first lay the chip
down on a non-sticky surface and re-grip it
before continuing.