Ch. 7. iDrive Imaging Sec. 7.3. Older Models
• Use only 0-80 x 7/64” Socket Head Cap Screws to attach the window assembly.
• Use only 0-80 x 1/8” Button Head Cap Screws to attach the cantilever holder clips.
• Use only 0-80 x 1/16” Cup Point Socket Set Screws for the piezo preload screw.
Due to wear and tear of use, the droplet holder accessory kit comes with replacement screws. Please
contact Asylum Research or your local Asylum distributor for additional hardware if proper replace-
ments cannot be obtained locally.
1.
Required tools and fasteners:
• 0.050” hex driver or Allen wrench for
the 0-80 x 7/64” socket head screws to
attach the window assembly.
• 0.035” hex driver or Allen wrench for
the 0-80 x 1/8” button head screws to
attach the cantilever holder clip.
• 0.028” hex driver or Allen wrench for
the 0-80 x 1/16” Cup Point Socket Set
Screws for the piezo preload screw.
Warning Using other fasteners than those
specified will damage your equipment.
2.
Loosen the piezo pre-load screw
Tools 0.028” hex driver or Allen wrench
• Loosen the piezo preload setscrew ¼
turn.
3.
Remove the spring clip
Tools 0.035” hex driver or Allen wrench
• Remove the screws securing the spring
clips to the droplet holder body.
• Remove the clip and the spacer plates
from the droplet holder body.
• Set the parts aside for cleaning.
BETA