37
Reviewing and Recording Measurement Data
Reviewing Measurement Results
The instrument can recall the last 500 measurement data.
1. Touch "DATA" on the home
screen.
2. On the history screen, the last 10 measurement data will be
displayed. To view the complete measurement history, use the
arrows pointing left and right to go back and forward a page
respectively.
All measurement results will be recalled once they are displayed.
The user-configured settings may be lost after repair or inspection. It is
recommended to keep a record of the details of each setting (see page 40 "Printing
SCALE Information" on how to print setting details).
Out of Memory Error
The instrument stores 500 measurement data. The instrument is
set at the factory to stop taking further measurements after the
500-measurement memory capacity is reached to avoid
accidental deletion of data (see page 61 "Error messages").
Uncheck the "DATA ERROR FULL" box to automatically
overwrite from the oldest measurement data when the memory
capacity is reached.