Instrument Specifications
(Common Specifications on Page 70 "Common specifications")
Automatic Digital Refractometer RX-5000i-Plus Cat.No.3275
Measurement range ① Refractive Index (nD): 1.32420 to 1.58000 ② Brix : 0.000 to 100.000%
Resolution ① Refractive Index (nD): 0.00001 ② Brix: 0.001%
③ Temperature: 0.01℃
Measurement accuracy
① Refractive Index (nD): ±0.00002 ② Brix: ±0.010%
(When measuring a standard sucrose solution of up to 50% Brix or standard
refractive index solution in MODE-1or MODE-T at 20℃)
Repeatability ① Refractive Index (nD): ±0.00001 ② Brix: ±0.010%
(When measuring a standard sucrose solution of up to 50% Brix or standard
refractive index solution in MODE-1or MODE-T at 20℃)
Dimensions and weight
37×26×14cm, 6.6kg (Main Unit only)
Automatic Digital Refractometer RX-5000i Cat.No.3276
Measurement range ① Refractive Index (nD): 1.32420 to 1.58000 ② Brix : 0.00 to 100.00%
Resolution ① Refractive Index (nD): 0.00001 ② Brix: 0.01% ③ Temperature: 0.01℃
Measurement accuracy
① Refractive Index (nD): ±0.00004 ② Brix: ±0.03%
(When measuring a standard sucrose solution of up to 50% Brix or standard
refractive index solution in MODE-1 at 20℃)
Repeatability ① Refractive Index (nD): ±0.00002 ② Brix: ±0.01%
(When measuring a standard sucrose solution of up to 50% Brix or standard
refractive index solution in MODE-1 at 20℃)
Dimensions and weight
37×26×14cm, 6.6kg (Main Unit only)
Automatic Digital Refractometer RX-7000i Cat.No.3279
Measurement range ① Refractive Index (nD): 1.29980 to 1.71500 ② Brix : 0.00 to 100.00%
Resolution ① Refractive Index (nD): 0.00001 (Factory default setting = 0.0001)
② Brix: 0.01% (Factory default setting =0.1%) ③ Temperature: 0.01℃
Measurement accuracy
① Refractive Index (nD): ±0.0001 ② Brix: ±0.1% *1
Repeatability ① Refractive Index (nD): ±0.00005 ② Brix: ±0.02% *1
Dimensions and weight
37×26×14cm, 7.0kg (Main Unit only)
*1 When measuring a standard sucrose solution of up to 50% Brix or standard refractive index solution in
MODE-1 at 20℃.
Guaranteed accuracy range is nD 1.32420 to 1.70000 and Temperature 5.0 to 70.0
°
C.
Automatic Digital Refractometer RX-9000i Cat.No.3278
Measurement range ① Refractive Index (nD): 1.29980 to 1.71500 ② Brix : 0.00 to 100.00%
Resolution ① Refractive Index (nD): 0.00001 ② Brix: 0.01% ③ Temperature: 0.01℃
Measurement accuracy
① Refractive Index (nD): ±0.00004
*2
② Brix: ±0.03%
*3
③ Brix: ±0.05%
*4
*1
Repeatability ① Refractive Index (nD): ±0.00002
*2
② Brix: ±0.01%
*3
③ Brix: ±0.01%
*4
*1
Dimensions and weight
37×26×14cm, 7.0kg (Main Unit only)
*1 When measuring a standard sucrose solution of up to 50% Brix or standard refractive index solution in
MODE-1. Guaranteed accuracy range is nD 1.32420 to 1.70000 and Temperature 5.0 to 70.0 °C.
*2 When nD 1.33299 to 1.42009, 10.00 to 30.00℃. For other range, nD ±0.00010 (Repeatability nD±
0.00005).
*3 When Brix 0.00 to 50.00%, 10.00 to 30.00℃.
*4 When Brix 50.01 to 95.00%, 10.00 to 30.00℃. For other range, Brix ±0.10%(Repeatability Brix ±
0.02%).