70
Common specifications
Measuring system Optical refraction critical angle detection system
Constant-temperature
unit
Built-in thermo-module
Measurement scales Refractive Index, Brix (Automatic Temperature Compensation),
User scale
Temperature
compensation range
5.00 to 75.00℃
Constant temperature
setting range
5.00 to 75.00℃ (No lower than 10℃ below or higher than 55℃ above the
ambient temperature)
Temperature display
accuracy
±0.05℃
Environmental
conditions
Use the instrument at an altitude below 2,000m (above sea level).
Use the instrument indoors.
Use the instrument where the temperature is between 5 to 40℃.
Use the instrument under the condition where humidity is 80% at 31℃ or
lower, falling linearly to 50% at 40℃.
Main supply voltage fluctuation should not to exceed ±10% the nominal
voltage.
Installation categories (Overvoltage Categories):ll
The pollution degree is 2 (according to IEC60664).
Display method 7.5-inch color LCD + touch screen
Display with top and
bottom limit bar
A visual indicator to approximate the measured value within a set range.
It can be turned off.
User scale When 3 to 5 points of a user scale corresponding to the Refractive Index are
inputted,
The internal software will automatically create the conversion formula.
Unit of measure (%, g/100g, g/100ml, mol/l, %max, %vol, no unit) can be
displayed.
Special scales 23 most commonly measured concentration scales are pre-programmed.
History function Up to 500 most recent measurements can be displayed and printed.
Printer output Digital printer DP-AD / DP-63(option)
Output method : RS-232C (9 pin D-Sub male connector)
Printed items: Refractive Index/Brix/user-scale, temperature at
measurement, year, month, day, time, measurement mode,
Sample No. and average value can also be printed.
Data output to USB
flash drive
CSV text file format
Communication with
computer
Communication via: USB-Serial
Output items: Refractive index, Brix, user scale, and temperature at
measurement
Input items: Zero-setting, selecting scales, and starting measurements
Zero adjustment Zero point is adjusted with purified water.
Manual calibration Indicated value can be changed within a certain range.
Light source LED (Approximating to D-Line wavelength)
Materials Prism: Artificial sapphire, Sample stage: SUS 316
Power supply AC100 to 240V 50/60Hz
Power consumption 90VA
When the number to the right of the terminating decimal place is four or less, it is rounded down. When
the number is five or more, it is rounded up.