Chapter 3 – Start-up and adjustments
UM-22100H-U User manual ATEQ D520 Page 38/122
2.3.6. Max fill
This function allows the setting of a maximum fill pressure which sets off an alarm if it is
exceeded.
) Adjust this parameter according to the method described in § 2.3.
2.3.7. Min fill
This function allows the setting of a minimum fill pressure threshold which sets of an
alarm if the pressure is not reached. This function is not active when an infinite test time
is set.
) Adjust this parameter according to the method described in § 2.3.
2.3.8. Reject unit
The reject units available depending on the system chosen by the operator are:
SI System: ml/h, l/h, ml/s, ml/min.
USA System: in
3
/h, ft
3
/h, in
3
/min, in
3
/h, ft
3
/h, in
3
/min.
Personalised system: no units, as chosen by the operator (CAL) only a unit name is
required, then the input of the maximum authorised drift. A learning cycle is then
necessary; it is carried out through the "CAL learning" special cycle.
) Adjust this parameter according to the method described in § 2.3.
2.3.9. Maximum reject
This parameter allows the definition of the upper limit of the authorised flow range in the
test part, above this limit the part is considered as bad.
) Adjust this parameter according to the method described in § 2.3.
2.3.10. Minimum reject
This parameter allows the definition of the lower limit of the authorised flow range in the
test part, under this limit the part is considered as bad.
) Adjust this parameter according to the method described in § 2.3.