Chapter 4 – Functions of the instrument
UM-22100H-U User manual ATEQ D520 Page 64/122
2.1.3. 23) Cycle end
This function allows the choosing of a different cycle end depending on the
configuration of the instrument (connection to a PLC or robot…).
a) Relay sequencing depending the End of Cycle
In order to interface the ATEQ D520 with its environment (PLC, PC …), the timing
charts below give the sequence for the electrical outputs (relay board on J8 connector)
and pneumatics (automatic connectors) according to the commands entered on the
front panel or on the J8 connector (START, RESET).
Legend
A
Coupling time for automatic connector A
B
Coupling time for automatic connector B
AZ
Piezo auto-zero time.
F
Fill time
S
Stabilization time
#
Undetermined time between the end of the programmed test time and
the pressing of the RESET key
T
Test time
D
Dump time
START
Press on the start key on the front panel or on pins 2-3 of the J3
connector.
RESET
Press on the reset key on the front panel or on pins 1-2 of the
J3 connector.
Automatic
connector
Active (high level): the pneumatic output is active (air exiting)
Inactive (low level): the pneumatic output is inactive (absence of air)
POK or PNOK
Good part or bad part relay on the J3 connector.
EoC
End of Cycle relay on the J3 connector
t mini
Minimum time for the taking into account of an input, 500 ms on the J8
connector of a central module and 50 ms on the J3 connector of an
instrument.
Actual times are not those displayed but those on the print-out.