Chapter 4 – Functions of the instrument
UM-22100H-U User manual ATEQ D520 Page 65/122
b) "Automatic RESET" cycle end
If the part is OK, at the beginning of the test time, the good part relay is activated until
the following test cycle. After the dump time, the end of cycle relay is activated (or after
the coupling time B if it is present).
If the part is fail, the fail part relay is activated as soon as the test is completed. The
instrument sends a cycle end signal. A new cycle can then be launched.
PRESSURE
START
RESET
A F S T
t
t
t
t
t
t
t
Program N°
Auto
connector A
t
Part good or
part bad
Time B
Time B
Auto
connector B
Cycle end
Test cycle
Control actions
t mini
t mini
t mini
Active
Disabled
Active
Disabled
D
The active program is the one selected before starting up. It remains active even if the
program inputs on the connector are no longer activated. This selection can only be
modified during the inter cycle period.
To return to program 1, when a cycle is not in progress, press any of the program
selection inputs.