Chapter 4 – Functions of the instrument
UM-22100H-U User manual ATEQ D520 Page 74/122
2.3.2. Sequences installation
First, check or validate the
SEQUENCE function in the
CONFIGURATION menu.
MENU
/
CONF
IGURATION
EXTENDED MENUS
AUTO SAVE : Yes
SEQUENCE : Yes
In the main menu, the
SEQUENCE function appears,
select it and validate it.
M
ENU
P
RINCIPAL
RUN SEQ. : SQ01
SEQUENCE
PARAMETERS
Select a first sequence, here
the sequence 01 then
validate.
S
E
QUE
NCE
UTILITY
SQ01
SQ02
In the line "NAME" it's
possible to type a sequence
name.
S
E
QUE
/SQ01
NAME :
<01> ----------
<02> ----------
Then select the first step and
validate, by using the UP and
DOWN arrows choose the
program and validate.
S
E
QUE
/SQ01
NAME :
<01> : Pr01+
<02> : ----------
Select the following step,
choose the second program,
validate and so on.
S
E
QUE
/SQ01
NAME :
<01> : Pr01+
<02> : Pr03+
When the sequence is
programmed, quit it by
pressing the C key.
C
S
E
QUE
/SQ01
NAME :
<01> : Pr01+
<02> : Pr03+
In the main menu, select the
desired sequence.
M
ENU
P
RINCIPAL
RUN SEQ. : SQ01
SEQUENCE
PARAMETERS
Start the sequence by
pressing the START CYCLE
key.
SQ01/
CYCLE/P
r:
001
PRESS = 30.0 mbar
READY
Note: if the sequence doesn't start or if it stops on a program test, check in this program
if the function and cycling conditions are validated.