56-101-01175 Revision H Page | 39
Figure 31 ATP Leak Test - Parameter Profile Screen
Figure 32 ATP Leak Test Results Screen
E. Go To Ground: This function will take the Test Set back to Ground (ambient barometric
pressure altitude) and CAS = 0 Kts. Go To Ground can be selected while operating in
Leak Measure or Control mode.
Note: During Power ON initializations, the Test Set measures ambient barometric
pressure “GROUND” and stores it in a temporary memory. When the user selects
Go To Ground, the Ps Target = ambient pressure altitude. After completion of Go
To Ground (Unit Safe at Ground) the Test Set measures ambient barometric
pressure again and that becomes the latest ambient barometric pressure altitude
at ground.
Main Menu: Go To Ground