9. Disinfection
Automatically set the uf rate to 3.6 l/h
Inhibit operation of the blood pump
Inhibit operation of the second blood pump
Automatically time the disinfection operation
Automatically infuse the correct amount of disinfectant
chemical into the system
Transition to a soft power off state, if selected by user
Provide the capability to take a manual blood pressure
reading, if the NIBP is installed
Resume the disinfection where that was running prior to
power loss (if power is cycled) if power is restored within
20 minutes.