3. Instrument Modes
• The heparin pump, dialysate flow rate, conductivity,
temperature, prescribed time, target UF, TMP and UF rate
functions cannot be accessed/changed.
• The UF rate is 3.60 L/h.
• The Event Timer is available.
For instruments with the Blood Pressure Monitor option:
• A manual blood pressure reading may be taken.
3.3.2.3 After Heat Disinfection
• The cabinet fan will operate as required.
• At the end of Cool Down (default selection), the Instrument
will enter the Rinse Mode once the temperature reaches
39.9°C.
• If Auto Off has been selected, the Instrument soft power is
automatically initiated at the end of Heat Disinfection.
3.3.3 Chemical Mode
3.3.3.1 Instrument status window displays
CHEMICAL
3.3.3.2 Instrument action
• Automated fluid path rinse, chemical infusion, dwell and rinse
appropriate for the selected chemical.
o Prompts operator to make necessary concentrate
line connections.
o Prerinse of fluid path until conductivity <1.0
mS/cm, then timed rinse.
o Prompts operator to make necessary concentrate
and disinfect line connections.
o Timed chemical infusion.
o Timed recirculation and/or dwell (as applicable for
chemical).
o Prompts operator to make necessary concentrate
and disinfect line connections.
o Timed forced rinse.
o Prompts operator to check for residual chemical.