3. Instrument Modes
The heparin pump, dialysate flow rate, conductivity,
temperature, prescribed time, target UF, TMP and UF rate
functions cannot be accessed/changed.
The UF rate is 3.60 L/h.
The Event Timer is available.
For instruments with the Blood Pressure Monitor option:
A manual blood pressure reading may be taken.
3.3.2.3 After Heat Disinfection
The cabinet fan will operate as required.
At the end of Cool Down (default selection), the Instrument
will enter the Rinse Mode once the temperature reaches
39.9°C.
If Auto Off has been selected, the Instrument soft power is
automatically initiated at the end of Heat Disinfection.
3.3.3 Chemical Mode
3.3.3.1 Instrument status window displays
CHEMICAL
3.3.3.2 Instrument action
Automated fluid path rinse, chemical infusion, dwell and rinse
appropriate for the selected chemical.
o Prompts operator to make necessary concentrate
line connections.
o Prerinse of fluid path until conductivity <1.0
mS/cm, then timed rinse.
o Prompts operator to make necessary concentrate
and disinfect line connections.
o Timed chemical infusion.
o Timed recirculation and/or dwell (as applicable for
chemical).
o Prompts operator to make necessary concentrate
and disinfect line connections.
o Timed forced rinse.
o Prompts operator to check for residual chemical.
157-1278-910 Rev A 3-11
November 2003