EasyManua.ls Logo

baxter ARENA - Recirculation

baxter ARENA
293 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9. Disinfection
It displays the message “Put Dialyzer Connectors on Rinse
Block” in the instruction window, if either sensor for the
dialysate connectors do not detect the presence of a connector.
It displays the message “Chem Disinfect: Infusion” in the
instruction window when the concentrate and dialysate
connectors are on the rinse block.
It sets the “A” concentrate pump rate according to Table 9-8.
It sets the “B” concentrate pump rate to the 20.13:1
proportioning ratio rate.
It cycles the bypass valve into bypass for 5 seconds every
minute.
It opens the dialysate filter purge valve #2 if Dialysate Filter
hardware is installed (refer to Section 24).
It enables the level control of the incoming water valve.
It de-energizes the recirculation valve.
It holds the elapsed time counter when any of the following
conditions occur:
o No Incoming Water alarm.
o Flow Restricted alarm.
o No Dialysate Flow alarm.
o Concentrate connectors are not on the rinse block.
o Dialysate connectors are not on the rinse block.
9.5.3 Recirculation
When a recirculation cycle is initiated, the Instrument will perform
the following actions.
It ensures the concentrate connectors are connected to their
assigned ports.
It displays the message “Chem Disinfect: Recirculation” in
the instruction window when the concentrate and dialysate
connectors are on the rinse block.
It sets the “A” concentrate pump rate to the 35:1
proportioning ratio rate.
It sets the “B” concentrate pump rate to thte 20.13:1
proportioning ratio rate.
It opens the rinse valve (dialysate pressure relief valve).
157-1278-916, Rev A 9-25
March 2004

Table of Contents

Related product manuals