Arena Service Manual
Handle used blood lines, dialyzers, transducer
protectors, and blood contaminated components as
potentially biohazardous waste. Refer to the clinic’s
procedures for the disposition of these items.
Make sure that the replacement of blood-contaminated
components is performed in an isolated area to help
prevent contamination of personnel and other devices.
Make sure that all tools used to handle, access, remove
and/or replace blood contaminated components are
disinfected when the job is done to prevent
contamination of personnel and/or other devices.
CAUTIONS
Static discharge may damage electronic components.
Both MOS and Bipolar integrated circuits may be damaged by
discharge of static electricity.
Both digital and linear integrated circuits may also be
damaged.
Static damage may occur not only to integrated circuits
mounted on the circuit board, but also to detached integrated
circuits.
Static damage may not be immediately evident.
Static damage is cumulative.
Do not remove or plug in any electronic component or
assembly while power is applied to the Instrument.
It is necessary to perform some of the adjustments with the
covers removed. Use extra care to prevent fluids from entering
or contacting the electronics of the Instrument.
Use proper ESD protection when handling or contacting any
ESD sensitive components.
Always repackage the removed ROMs or boards in the
original ESD packaging.
Perform work using standard ESD precautions including
personal grounding strap.
16.8.2 Contamination Assessment Procedure
If during the dialysis treatment the external transducer protector is
wetted, and blood is visible on the Instrument side of the
transducer protector, then the Instrument must be removed from
service at the end of treatment, and a qualified technician must
assess the Instrument for contamination using the following
16-14 157-1278-923 Rev A
February 2004