Arena Service Manual
9.6.2.4 Procedure
1. Make sure that there is an adequate air gap between the drain
line and the drain.
If the type of chemical disinfectant is changed, rinse the
disinfect line with water for approximately 2 minutes before
connecting the line to the container of disinfectant.
2. Connect the disinfect line (yellow connector) to a container of
1 liter of household bleach.
3. Connect the acid/acetate concentrate line (pink connector) to
the disinfect port (yellow).
4. Allow bleach to infuse into the fluid path.
5. After the 60 minutes, visually check the drain line for deposits.
If the drain line is clean, disconnect the Instrument from the
bleach supply.
If the drain line is not clean, continue to infuse bleach until
the line is clean (add more bleach to the container as
needed).
6. To disconnect the Instrument:
a. Connect the acid/acetate concentrate line (pink connector)
to the acid/acetate rinse port (pink).
b. Wait approximately 15 seconds for the disinfect line to
drain, then connect the disinfect line (yellow connector) to
the disinfect port (yellow).
CAUTION
Do not allow sodium hypochlorite to remain in the fluid path
longer than the recommended time or damage may result.
7. Rinse the Instrument for approximately 30 minutes. Continue
rinsing until a test specific for the presence of sodium
hypochlorite is negative.
8. Repeat the procedure as needed to remove protein buildup
from the drain line.
9. Make sure that there is an adequate air gap between the drain
line and the drain.
9-30 157-1278-916, Rev A
March 2004