Arena Service Manual
approximately 90 seconds after the blood pump is turned on,
the blood pump rate is changed or the UF rate is changed, or
the RESET button touched.
The minimum low venous pressure alarm limit is
approximately +10 mmHg.
The air and blood leak detector instrument alarm responses are
active.
If the extracorporeal alarms are disarmed:
The arterial pressure alarm limits are –300 and +600 mmHg.
The venous pressure alarm limits are ±200 mmHg around the
indicated venous pressure when the ARMED / DISARM
button is pressed or 10 seconds after the blood pump starts,
blood pump power switch is turned off or the blood pump rate
is changed, or the level adjust button pressed.
The TMP alarm limits are ±200 mmHg (within the range of the
technician set minimum TMP value to +500 mmHg) around
the indicated TMP approximately 90 seconds after the UF rate
is changed or the blood pump is turned on, off or rate changed,
or the RESET button touched.
The air and blood leak detector instrument alarm responses are
disabled (except for the visual indicator).
For instruments with the Blood Pressure Monitor option:
A manual blood pressure reading may be taken.
The blood pressure schedule may be set, changed or manually
stopped.
3.2.2 Dialyze Mode
3.2.2.1 Instrument status window displays
DIALYZE
3.2.2.2 Instrument action
All alarms are functional. No alarm condition exists.
The blood pump will operate.
The venous line clamp is open.
The UF rate is at the calculated value (from the PRESCRIBED
TIME and TARGET UF), unless manually overridden or the
calculated UF rate is less than the technician set minimum rate.
The heparin pump will operate.
3-6 157-1278-910 Rev A