3. Instrument Modes
• The elapsed time is accumulated.
• While the blood pump is stopped and/or the Instrument is in
bypass, the elapsed time is not accumulated.
• The main alarm lamp is off.
• The accumulated ultrafiltrate removed is displayed.
• The TMP alarm limits are set to ±35 mmHg (within the range
of the technician set minimum TMP value to +500 mmHg)
around the indicated pressure (reference pressure),
approximately 90 seconds after the Dialyze Mode is started or
the UF rate is changed, blood pump is turned on, the blood
pump rate is changed, or the RESET button is touched.
• The TMP alarm limits are spread to a maximum 600 mmHg
span (585 mmHg default) around the reference pressure for
approximately 90 seconds when the blood pump or the UF rate
is changed, then sets to a new ±35 mmHg window.
• The maximum high TMP alarm limit is +500 mmHg. The
minimum low TMP is set by the technician.
• The SET LIMITS button may be used to manually set the
TMP, arterial pressure, venous pressure alarm limit windows
around the appropriate indicated pressure.
• The arterial pressure alarm limits spread to –300 and +600
mmHg, for approximately 10 seconds, when the blood pump is
started or rate is changed, or the level adjust button pushed, or
the RESET button touched.
• The venous pressure alarm limits spread to ±200 mmHg
around the indicated venous pressure, for approximately 10
seconds, when the blood pump is started or rate is changed, or
the level adjust button pushed, or the RESET button touched.
• The arterial and venous pressure alarm limits set to ±50 mmHg
of the indicated value approximately 10 seconds after the blood
pump is started or rate is changed, the level adjust button
pushed, or the RESET button touched.
• The minimum low venous pressure alarm limit is
approximately +10 mmHg.
• The main alarm lamp flashes and the audible alarm beeps 3
times approximately every 90 seconds, when the Instrument is
in manual bypass or minimum UF, or the blood pump is off.
• The Event Timer is available.