Arena Service Manual
5.5 BLOOD PUMP POWER BOARD
This board is controlled directly by the Blood Pump Controller
board. It is located right behind and above the blood pumps. The
BP Power board in turn controls the operation of the analog
devices related to the blood pump system (see Figure 5-1):
Blood Pumps
Heparin Pump
Level Adjust Motors, valves and switches
Venous and Arterial Membrane switches
Optional Arterial Line Clamp
Pressure Transducers
Speaker power
Fan
Sonalert
5.6 UF-PROPORTIONING POWER BOARD
This board is controlled directly by the UF- Proportioning
Controller board which incorporates buffering for the gear pumps.
The UF- Proportioning Power board in turn controls the operation
of some of the analog devices related mostly to the UF-
Proportioning system and fluid path (see Figure 5-1):
Supply Pump
Dialysate Pressure Pump
Deaeration Pressure Pump
End-of-Stroke Sensors
Heater Control
UF Flow meter valves
Flow Equalizer Valves
A & B Proportioning Pumps
Dialysate Pressure Relief Valve
Conductivity-Temperature measurement
OLHDF Purge Valve # 2 (optional)
Citric Acid Valve
5-30 157-1278-912 Rev A