9. Disinfection
Make sure that the determination test is specific for the
chemical used.
NOTE
An error message will appear on the screen accompanied by an
audio “beep” while the acid connector is removed from the
machine. This error message states, “Put concentrate connectors
on block”. Mute this error while infusing citric acid.
When the citric acid has been infused, replace the acid connector
on the machine.
9.3.5 Integrated Heat Clean
The Integrated Heat Clean Disinfection is a future option that will
permit the operator to preset the Instrument to automatically start
an Integrated Heat Clean disinfection cycle at a preset time and
infuse hot water from a heated water treatment system, thus heat-
disinfecting the Instrument’s incoming water line, fluid path, and
drain line.
This option will require hardware changes of the following
elements capable of withstanding the heat:
Inlet water line
Drain line
Inlet water pressure regulator.
See Sections 17.3.5.1, 17.3.5.2, and 17.3.5.5 for customizing
information.
157-1278-916, Rev A 9-13