Arena Service Manual
Main alarm lamp is off (except for operator prompts and error
situations).
Audio alarm is muted (except for operator prompts and error
situations).
The blood pump will not operate.
The heparin pump, dialysate flow rate, conductivity,
temperature, prescribed time, target UF, TMP and UF rate
functions cannot be accessed/changed.
The UF rate is automatically set to 3.6 L/h.
ELAPSED TIME display indicates the elapsed time of each
disinfection phase or HOLD as appropriate.
The operator may select from technician set chemical list.
Formaldehyde dilution is dependent upon technician setting.
Auto soft power off during dwell for some chemicals.
If dwell time is not complete when the Instrument is powered
up, the operator is prompted about the remaining time and the
Instrument automatically soft powers off.
The operator may manually start the forced rinse.
Forced rinse flow rate automatically set to technician set value.
Forced rinse time dependent upon technician set residual
chemical sample location, chemical and forced rinse flow rate.
The Event Timer is available.
For instruments with the Blood Pressure Monitor option:
A manual blood pressure reading may be taken.
3.3.4 Shut Down
3.3.4.1 Instrument status window displays
SHUTDOWN
3.3.4.2 Instrument action
The blood pump is stopped.
The venous line clamp is clamped.
All pumps are stopped.
UF flow meter is turned off.
Audio alarm sounds.
3-12 157-1278-910 Rev A
November 2003