Arena Service Manual
Perform work using standard ESD precautions including
personal grounding strap.
Before disconnecting any cables, note their position. Mark
as needed.
4. Disconnect the other end of the contaminated tubing from the
filter(s).
5. If the filter is contaminated, replace it.
Carefully inspect the entire tubing set for blood. Replace all
contaminated components. Since you can not inspect the
interior of the transducer manifold, solenoid valve and valve
manifold (as applicable), these components may need to be
replaced if the filter is contaminated.
6. Connect new tubing between the filter(s) and the pressure
fitting(s).
16-16 157-1278-923 Rev A
February 2004