Arena Service Manual
Note
When DIP Switch 1 is ON, it allows the Instrument to enter
Service/Calibration Mode. The external Reed Service Switch is
connected in parallel to this switch and can be used to
momentarily turn on Switch 1.
6. UF Profile: space allocated for six profiles, each consisting
of up to six hours of treatment data in fifteen minute
intervals. Each profile also contains four bytes for setting
UF Only treatment in fifteen minute intervals
7. Sodium Profile: space allocated for six profiles, each
consisting of up to six hours of sodium concentrate level
dispensing in fifteen minute intervals
8. Bicarbonate Profile: space allocated for six profiles, each
consisting of up to six hours of bicarbonate concentrate
level dispensing in fifteen minute intervals
9. Blood Pressure Data
• Systolic Minimum Limit
• Diastolic Minimum Limit
• Systolic Maximum Limit
• Diastolic Maximum Limit
• Storage of Associated Time
• Systolic Pressure
• Diastolic Pressure
• Pulse Rate
• Mean Arterial Pressure
• Time
• Type
• Scheduled / Manual
• Sitting / Standing
• Capacity for up to 49 Blood Pressure Records
10. Treatment Setup Recall
• Storage of previous treatment parameters
• Prescribed Time
• Heparin Rate
• Heparin Auto Shutoff
• BP Schedule Interval
5-28 157-1278-912 Rev A
November 2003