Arena Service Manual
Note
When DIP Switch 1 is ON, it allows the Instrument to enter
Service/Calibration Mode. The external Reed Service Switch is
connected in parallel to this switch and can be used to
momentarily turn on Switch 1.
6. UF Profile: space allocated for six profiles, each consisting
of up to six hours of treatment data in fifteen minute
intervals. Each profile also contains four bytes for setting
UF Only treatment in fifteen minute intervals
7. Sodium Profile: space allocated for six profiles, each
consisting of up to six hours of sodium concentrate level
dispensing in fifteen minute intervals
8. Bicarbonate Profile: space allocated for six profiles, each
consisting of up to six hours of bicarbonate concentrate
level dispensing in fifteen minute intervals
9. Blood Pressure Data
Systolic Minimum Limit
Diastolic Minimum Limit
Systolic Maximum Limit
Diastolic Maximum Limit
Storage of Associated Time
Systolic Pressure
Diastolic Pressure
Pulse Rate
Mean Arterial Pressure
Time
Type
Scheduled / Manual
Sitting / Standing
Capacity for up to 49 Blood Pressure Records
10. Treatment Setup Recall
Storage of previous treatment parameters
Prescribed Time
Heparin Rate
Heparin Auto Shutoff
BP Schedule Interval
5-28 157-1278-912 Rev A
November 2003