EasyManua.ls Logo

BAYKON MX08 - N-Bus Addressing Via Xface

BAYKON MX08
148 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MX08 Multichannel Weighing System, Technical Manual, Rev. 2.5 Page 22 of 144
4.3. N-Bus Addressing via xFace
N-Bus Addresses of Analog Digitizer Instrument.
Read Button: Reads the Address and N-MAC Numbers of Analog Digitizers.
Write Button:
Writes the N-Bus address to Analog Digitizers related to N-MAC number.
Auto Address: Fills the address fields from 0 to 7.
N-Bus Addresses of I/O Instrument.
Read Button:
Reads the Address and N-MAC numbers of I/O units.
Write Button: Writes the N-Bus address to I/O units related to N-MAC number.
Auto Address:
Fills the address fields from 0 to 7.
Figure 3.7 - N-Bus Addressing tab
For the first time connection to MX08 system, only Gateway and Addressing tabs becomes
active. First, address the instruments to N-Bus to activate the instruments in MX08 system.
Addressing Analog Digitizers to N-Bus: First, click ‘Readbutton to read the N-MAC
numbers and the addresses. Give an address number to each Analog Digitizers. Then
click Writebutton to save the addressing settings. If the addressing Analog Digitizers to
N-Bus is completed successfully; Setup, Calibration and Digitizer Status tabs, Scale
Address and Visual Weight Display will be activated.
Addressing Input/Output Instruments to N-Bus: First, click Read button to read N-
MAC numbers and addresses. Give an address to each MX08 IO. Then click ‘Write button
to save the addressing settings. If the addressing MX08 IOs to N-Bus is completed
successfully, I/O Status tab will be activated.
Note: Sequentional address numbers of N-Bus slave instruments are recommended.

Table of Contents

Related product manuals