MX08 Multichannel Weighing System, Technical Manual, Rev. 2.5 Page 73 of 144
9.5. Profibus DP Data Structure
Please refer to the Section 3.5 for detailed PLC configuration.
MX08 PB Output to PLC Input
Bitwise of Dword:
1st
Dword
(R)
Active MX08 IO instruments
Active MX08 AD instruments
Error Table
1
st
Dword Description
0: No instrument found 1: 7 addressed MX08 IO is active
instruments
0: No instrument found 1: 6 addressed MX08 IO is active
0: No instrument found 1: 5 addressed MX08 IO is active
0: No instrument found 1: 4 addressed MX08 IO is active
0: No instrument found 1: 3 addressed MX08 IO is active
0: No instrument found 1: 2 addressed MX08 IO is active
0: No instrument found 1: 1 addressed MX08 IO is active
0: No instrument found 1: 0 addressed MX08 IO is active
0: No instrument found 1: 7 addressed MX08 AD is active
instruments
0: No instrument found 1: 6 addressed MX08 AD is active
0: No instrument found 1: 5 addressed MX08 AD is active
0: No instrument found 1: 4 addressed MX08 AD is active
0: No instrument found 1: 3 addressed MX08 AD is active
0: No instrument found 1: 2 addressed MX08 AD is active
0: No instrument found 1: 1 addressed MX08 AD is active
0: No instrument found 1: 0 addressed MX08 AD is active
D15
D14
D13
D12
of MX08 PB
0001
System fails.
− Re-energize the instrument. If seen again, change the board.
0010
EEPROM fail.
− Re-energize the instrument. If seen again, change the board.
0011
− Re-energize the instrument.
− Check the field-bus module is installed on the board.
− If seen again, change the board.
0100
No Instrument found on N-Bus.
− Install Digitizer or I/O instruments to the system.
− Check N-BUS connectors whether they are installed.
0101
An instrument is installed/removed in/from system.
− Re-address the instruments on N-Bus (Refer to section 3.3
or
Section 3.9 N-Bus addressing)
The command is applied successfully