MX08 Multichannel Weighing System, Technical Manual, Rev. 2.5 Page 47 of 144
6.3. Error Table
The MX08 instruments have been designed as very reliable and virtually error free
instruments. However if an error occurs do not attempt to repair the equipment before you
understand what caused the error. Note the problems you have with your instrument and
the error messages shown by the LEDs located on the front panel. Then try to solve the
problem according to the error table given below.
Description Actions to take / Possible cause
- - - - -
N-Bus
Communication
Error
− Re-address the instruments on N-Bus (Refer to
section 3.3 or Section 3.9 N-Bus addressing).
−
Check the Gateway instrument whether it is in
system and powered on.
− Check the N-Bus connector wheth
er they are
installed on the rail.
PROG Programming − MX08 is in programming mode.
SYS System Error
− Re-
energize the instruments. If seen again, change
the board of related instrument.
ADC ADC Error
− Check the load on MX08 AD.
− Check the calibration of MX08 AD.
− Load cell or instrument could be defective.
− Re-
energize the instrument. If seen again, change
the board.
OVER Over load
− Check the load of MX08 AD.
− Check the calibration of MX08 AD.
− Load cell or instrument could be defective.
UNDER Weight is too low
Table 5.2 - Error table