MX08 Multichannel Weighing System, Technical Manual, Rev. 2.5 Page 19 of 144
3.6. Commissioning
PRECAUTION: Please read this manual carefully before energizing the instrument.
Perform the commissioning operation according the procedure given in this section. Only
trained person is allowed for cleaning, commissioning, checking and servicing of the
instrument. The interference of untrained person may cause some unwanted damages or
injures.
Before power on the instrument, please make the required mechanical and electrical
installations. After power on, you have to program your MX08 system before field bus
interfacing.
Install xFace to your PC as described in Section 3 Programming. xFace software is used
for programming, calibration and testing of MX08 system.
After checking the performance of instruments with xFace, you can begin to use MX08
system in your application.
4. PROGRAMMING
PRECAUTION: Please read this manual carefully before energizing the instrument.
Perform the commissioning operation according the procedure given in Section 2.6. Only
trained person is allowed for cleaning, commissioning, checking and servicing of the
instrument. The interference of untrained person may cause some unwanted damages or
injures.
MX08 system needs to be programmed in sequence below before field bus interfacing;
• Installing xFace software to your PC.
• Connection between PC and MX08 gateway
• N-Bus addressing
• Gateway set up
• Analog Digitizer programming and calibration
• Analog Digitizer performance tests
• Digital Input / Output tests
MX08 family instruments are programmed and calibrated by xFace software given with the
instrument. The latest version of xFace software can be downloaded from BAYKON
website www.baykon.com.
4.1. Installation of xFace
Please follow following steps to install the xFace software:
• Close all applications on your PC.
• Insert the CD that contains the xFace software into the CD-ROM drive.
• Double click to setup.exe to start installation. The setup Wizard is displayed.