MX08 Multichannel Weighing System, Technical Manual, Rev. 2.5 Page 28 of 144
Address number is [7]. Increases address number to [0].
Description of LED Description of Programming Switch
Its N-Bus address number is
same as gateway’s selected
address number.
As
signs gateway’s selected address
number to its N-Bus address number.
Its N-Bus address number is
not same as gateway’s
selected address number.
Note: Sequentional address numbers of N-Bus slave instruments are recommended.
4.10. Back-up Parameters and Calibration
The parameter settings and calibration coefficients can be saved to the back up file after
reading these data from MX08 instruments. This back up file can be written to MX08
instruments after opening it by xFace.
Back up feature gives service advantage to MX08 instruments. Proposal is burning these
files to CD/DVD and saving this back up CD/DVD in the control cabinet additional saving
file in your PC.
Important note: The previous parameter values and calibration coefficients in MX08 AD
will be lost after loading back up to the instrument.